The PLC Tools SIM-ALD Current Loop Simulator Dual Channel simulates two independent 2-wire loop powered transmitter channels.
For each channel five fixed output settings are available: 4 mA, 8 mA, 12 mA, 16 mA or 20 mA.
A specific output selection is made by connecting signal wire to one of the five output terminals.
It is not recommended to connect more than one output at the same time. If this happened, maximum current limited to 24mA per channel.
Simulator can work with wide range of loads from 0 to 1000 Ohm, typical analog input impedance is 250 Ohm. Loads above 500 Ohm require 24V or higher loop power supply. A typical 2-wire installation requires a loop power supply usually provided by an external 24VDC power supply.
Some sourcing analog input modules can provide loop power also. Follow the analog input module specification for a power supply selection.
Typical Two-Wire Simulator Connection with an External Loop Power Supply
Typical Two-Wire Simulator Connection with a Sourcing Analog Input
Important:
Do not use this device for calibration of input modules or instruments. This device is a simple tester that provides current signal within a selected range.
Do Not Operate Device Unless Area Is Non-Hazardous
Specifications:
Device Simulation. Two-Wire Loop Powered Devices
Power Supply. 12-30V DC 24VDC typical
Current Output
Number of Channels: 2 Isolated
Output Range: 4mA to 20mA
Fixed Preset: 4 , 8, 12, 16, 20 mA
Output Selection: Terminal
Output Accuracy: ±1% of full scale( 0.2mA)at 25°C
Protection: Reverse Polarity
Load Range: 0-1000 Ohm 250 Ohm typical
Loads over 500 Ohm require 24V or higher power supply
Currently the device does not carry any agency approvals and is not compliant with RoHS.
Disclaimer
This device is intended to provide general assistance with current loop debugging, testing and application development.
It should not be permanently used in live production systems.
Accordingly, production system must be tested and commissioned with real instruments to ensure safe and reliable operation.
IN NO EVENT SHALL THE DEVICE MANUFACTURER BE LIABLE FOR ANY DAMAGES OF ANY KIND INCLUDING DIRECT, INDIRECT, INCIDENTAL,CONSEQUENTIAL, LOSS OF PROFIT OR DAMAGE.
The examples and diagrams in this manual are included for illustrative purposes only. Because of the many variables and requirements associated with any particular installation, the device manufacturer cannot assume responsibility or liability for actual use based on the examples and diagrams.
Before making any decision or taking any action that might affect your equipment, you should consult a qualified professional advisor.